Electric Clean Steam Generators

Lautenschläger electric steam generators are specifically designed to meet high purity requirements and to serve applications with high peak steam consumption (e.g., steam sterilizers, steam disinfectors, and steam pasteurizers). A large boiler volume, high operating pressure, and low heating surface load ensure optimally dry saturated steam.

Elektrodampferzeuger Reindampferzeuger Dampferzeuger

Lautenschläger Electric Steam Generators for clean saturated steam

  • Pressure Equipment Directive (PED) 2014/68/EU category II or III, depending on boiler size

  • PLC control with color touch panel

  • Optionally with energy / steam management system and peak-load management to ensure an optimal supply to multiple large consumers

  • Equipment options such as thermal degassing of the feedwater, conductivity-controlled desalination, automatic sludge removal, and condensate recovery

  • Horizontal boilers with a large water surface area for optimally conditioned saturated steam

Technical Data

Type heating Heating Capacity [kW] Ø Capacity [kg/h] Peak Capacity at max. [bar]
ED 15 electric 15 20 1 kg / 1 min 5
ED 36 electric 36 48 2 kg / 1 min 4
ED 72 electric 72 96 5 kg / 1 min 5
ED 144 electric 144 194 12 kg / 2 min 4
ED 144 S electric 144 194 12 kg / 2 min 10

Pure steam supply lines for steam sterilization

German

English

When designing a pure steam supply system, the requirements are entirely different from those for a process or heating steam supply. In addition to material selection and purity requirements, the process characteristics of the steam consumers often result in additional specifications. For example, pure steam for autoclaves in laboratories or in industrial and pharmaceutical applications must not be superheated, nor may it contain excessive moisture. Furthermore, it should be as free as possible of non-condensable gases. Our brochure “Pure steam supply lines for steam sterilization“ was written with the aim of describing the requirements for pure steam supply systems in steam sterilization technology and identifying possible technical solutions.

Steam used in sterilization processes should be saturated steam that is free of impurities and contains no non-condensable gases. Impurities in the sterilization steam enter the autoclave and are transferred to the products. This can damage or contaminate your products, cause unwanted deposits to form, or even endanger human life. These residues in the steam can originate from the feedwater, the steam boiler, or the steam lines. Our brochure provides many valuable tips for optimizing your steam supply system.